Improvements to the absolute photometric calibration of IRAC

19 Apr 2012, 14:50
20m
WH1W

WH1W

Paper Session 4C

Speaker

Dr Sean Carey (Spitzer Science Center / IPAC)

Description

We have made several significant improvements to the final cryogenic calibration of the IRAC instrument aboard the Spitzer Space Telescope. The final calibration uses the ensemble of calibration data collected over the five-plus years of the cryogenic mission and includes knowledge obtained on intrinsic photometric variations. Photometric variation with location on the array and phase of stellar centroid with respect to pixel center have been characterized and corrected for. The stability of the instrument and large amount of calibration data available permit correction of these effects to better than a percent. The uncertainty in the final calibration is dominated by the uncertainty in calibrating the fundamental photometric calibrators used to provide the zero point of the IRAC photometric system. Planned observations of the calibrators Sirius and 109 Vir will be discussed as well as the current results of several cross-calibration experiments with WISE and HST. This work is based on observations made with the Spitzer Space Telescope, which is operated by the Jet Propulsion Laboratory, California Institute of Technology under a contract with NASA. Support for this work was provided by NASA through an award issued by JPL/Caltech

Primary author

Dr Sean Carey (Spitzer Science Center / IPAC)